Skip to main content
← Signals and displays
Simulator reference

EEG suppression ratio

Fraction of 0.25 s sub-epochs under 5 uV peak-to-peak in a 60 s window, runs under 1 s ignored. Code file is outside multimodal/.

Sources the model was built from

  1. Hirsch LJ, Fong MWK, Leitinger M, et al. American Clinical Neurophysiology Society's Standardized Critical Care EEG Terminology: 2021 Version. J Clin Neurophysiol. 2021;38(1):1-29.

Related literature

Each paper was checked against this feature by independent automated reviewers before being listed. The reviewers worked from findings extracted from the paper; a clinician has not read each one.

  1. Connor CW. OpenBSR: An Open Algorithm for Burst Suppression Rate Concordant with the BIS Monitor. Anesth Analg. 2025;140(1):220-223.
  2. Lacan L, Betrouni N, Lamblin MD, et al. Quantitative approach to early neonatal EEG visual analysis in hypoxic-ischemic encephalopathy severity: Bridging the gap between eyes and machine. Neurophysiol Clin. 2021;51(2):121-131.
  3. Jegarl AM, Walline MC, Goldstein PA, et al. Is Electroencephalographic Burst-Suppression Good, Bad, or Indifferent for Brain Health? Context Matters. Anesth Analg. 2026;142(6):1169-1179.
  4. Yuan I, Bong CL, Chao JY. Intraoperative pediatric electroencephalography monitoring: an updated review. Korean J Anesthesiol. 2024;77(3):289-305.
  5. Veciana de Las Heras M, Sala-Padro J, Pedro-Perez J, et al. Utility of Quantitative EEG in Neurological Emergencies and ICU Clinical Practice. Brain Sci. 2024;14(9).
  6. Chakravarty S, Baum TE, An J, et al. A hidden semi-Markov model for estimating burst suppression EEG. Annu Int Conf IEEE Eng Med Biol Soc. 2019;2019:7076-7079.
  7. Westover MB, Ching S, Kumaraswamy VM, et al. The human burst suppression electroencephalogram of deep hypothermia. Clin Neurophysiol. 2015;126(10):1901-1914.
  8. Purdon PL, Sampson A, Pavone KJ, et al. Clinical Electroencephalography for Anesthesiologists: Part I: Background and Basic Signatures. Anesthesiology. 2015;123(4):937-60.
  9. Bao H, Wang J, Cui Z, et al. A Systematic Review of How Cardiopulmonary Bypass Parameters Influence Electroencephalogram Signals. Brain Sci. 2026;16(4).
  10. van Putten MJAM, Hofmeijer J. Neurophysiologic Mechanisms of Cerebral Ischemia and the EEG. J Clin Neurophysiol. 2026;43(5):375-383.
  11. Manohara N, Ferrari A, Greenblatt A, et al. Electroencephalogram monitoring during anesthesia and critical care: a guide for the clinician. J Clin Monit Comput. 2025;39(2):315-348.
  12. Bögli SY, Cherchi MS, Beqiri E, et al. Association between EEG metrics and continuous cerebrovascular autoregulation assessment: a scoping review. Br J Anaesth. 2024;133(3):550-564.
All 20 papers
  1. Fleischmann A, Georgii MT, Schuessler J, et al. Always Assess the Raw Electroencephalogram: Why Automated Burst Suppression Detection May Not Detect All Episodes. Anesth Analg. 2023;136(2):346-354.
  2. Benedetti GM, Guerriero RM, Press CA. Review of Noninvasive Neuromonitoring Modalities in Children II: EEG, qEEG. Neurocrit Care. 2023;39(3):618-638.
  3. Nguyen G, Postnova S. Progress in modelling of brain dynamics during anaesthesia and the role of sleep-wake circuitry. Biochem Pharmacol. 2021;191:114388.
  4. Shanker A, Abel JH, Schamberg G, et al. Etiology of Burst Suppression EEG Patterns. Front Psychol. 2021;12:673529.
  5. Narula G, Haeberlin M, Balsiger J, et al. Detection of EEG burst-suppression in neurocritical care patients using an unsupervised machine learning algorithm. Clin Neurophysiol. 2021;132(10):2485-2492.
  6. Mullaguri N, Beary JM, Newey CR. Early detection of brainstem herniation using electroencephalography monitoring - case report. BMC Neurol. 2020;20(1):406.
  7. Bojak I, Stoyanov ZV, Liley DT. Emergence of spatially heterogeneous burst suppression in a neural field model of electrocortical activity. Front Syst Neurosci. 2015;9:18.
  8. Kenny JD, Westover MB, Ching S, et al. Propofol and sevoflurane induce distinct burst suppression patterns in rats. Front Syst Neurosci. 2014;8:237.

Editorial draft · clinical review pending. A teaching model, not calibrated to an individual child.